Keysight N5990A MIPI M-PHY Receiver Test User Guide
177
HS Tests
5
Procedure
The data generator sends Burst-mode CJTPAT signaling while the DUT is
configured for Terminated mode. The differential voltage amplitude and
the common-mode are set to the nominal value (200mV). No other signal
impairments such as the ISI channel or jitter sources are added to the
signal.
A sweep of the calibrated Lane-to-Lane skew is done in the tested lane,
starting with the minimum skew value (Min Tested Value) and increasing
with the given Step Size value until the Max Tested Value is reached. At
each step, the BER is measured and compared with the Target BER to
consider the pass/fail result.
This test is performed for all selected HS lanes and gears and only for
Terminated case.
Since for JBERT
N4903B
, it is not possible to skew the generator outputs,
the lane to lane skew is performed by pattern. So the skew resolution is
limited to 1 UI.
Connection Diagram
show the connection diagrams for the different
system configurations.
NOTE
The connection diagrams shown are for Gear1. They are the same for
Gear2 and Gear3 but vary in the TTC value (G1: 250ps; G2: 125ps; G3
and G4: 60ps).
Содержание N5990A
Страница 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Страница 17: ...Keysight N5990A MIPI M PHY Receiver Test User Guide 1 Introduction Document History 18 Overview 19...
Страница 62: ......
Страница 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 214: ......
Страница 234: ......
Страница 268: ......
Страница 298: ......