Keysight N5990A MIPI M-PHY Receiver Test User Guide
77
Calibrations and Test Procedures
3
Eye Width Max. Variation
It is the allowed variation of the target eye width.
Eye Height Target
It is the eye height to be calibrated.
Filter Delay Chan1/2
The filter delay can be enabled on the oscilloscope.
Frequency Deviation
It is the deviation that has occurred in the frequency value.
Frequency Range
It is the range of the jitter frequency values to be tested. It is defined with four values separated by semicolons:
<Start Value>;<End Value>;<Number of Steps>;<Scale Type>. The scale type can be Linear or Logarithmic.
HS Sequence File
It is the sequence file to be used during the calibration/test.
HS Sync length
It is the number of sync pattern symbols before the data burst. It must be specified in SI (1SI = 10UI).
Initial Jitter Amplitude Step
It is the initial step size value of jitter amplitude.
IBerReader Init Mode
It refers to the configuration parameters of the BER reader for the system initialization mode with the
parameters such as data channel, data rate and termination mode (Terminated / Unterminated).
ISI (p-p)
It is the amount of ISI (Inter Symbol Interface) added during the test procedure.
ISI trace / J20
It is the amount of ISI (Inter Symbol Interface) added during the test procedure. It is only available for J-BERT
configurations with J20 option.
Jitter Calibration File
It is the sequence file for the jitter calibration.
Jitter Frequencies
These are the frequency points to be tested/calibrated.
Jitter Frequency Range
It is the range of the jitter frequency values to be tested. It is defined with four values separated by semicolons:
<Start Value>;<End Value>;<Number of Steps>;<Scale Type>. The scale type can be Linear or Logarithmic.
Jitter Amplitude Min Step
It is the minimum step size value that is used to find the DUT limit for each jitter frequency.
Jitter Increase Accuracy
This parameter is available in the Jitter Tolerance test (2.1.7) when the 'Perform Jitter Limit Test" property is
set to true. It defines the windows size of the binary search that is used to find the maximum tolerated jitter.
Lane Under Test Termination Model
For M8020A, system configuration is necessary to indicate the termination model such as 50Ohm or 100Ohm.
If the selected mode is not right, the M8020A outputs will be turned off. When the 100 Ohm Termination board
is not available, you can set this parameter to “50 Ohm” and connect SMA cables directly to the scope.
Level Pairs
They are pairs of single ended amplitudes (differential and common mode voltage) separated by “|”.
LFSJ Frequencies
These are the frequencies to be tested during Jitter Tolerance Test.
Max. Calibration Voltage Amplitude
It is the start voltage value to find the target eye height.
Max Interference Amplitude
It is the maximum value of the interference amplitude to be tested.
Max Interference Value
It is the higher limit for the tested RF (Radio Frequency) Interference. If the minimum and maximum
interference values are equal, only one interference level is tested.
Max Jitter Value
It is the maximum jitter value to be calibrated.
Parameter Name
Parameter Description
Содержание N5990A
Страница 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Страница 17: ...Keysight N5990A MIPI M PHY Receiver Test User Guide 1 Introduction Document History 18 Overview 19...
Страница 62: ......
Страница 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 214: ......
Страница 234: ......
Страница 268: ......
Страница 298: ......