Keysight N5990A MIPI M-PHY Receiver Test User Guide
139
Calibrations
4
Procedure
The N4903B does a sweep of the ISI traces starting with the Min Jitter
Value and increases until the Max Jitter Value is reached. At each step, the
actual ISI amplitude is measured with the oscilloscope and is stored along
with the set ISI trace.
The Max Jitter Value should be decreased as the Gear increases, otherwise
the calibration fails and the log file will show a warning message as: "Jitter
can not be measured".
Connection Diagram
Refer to
for the connection diagram.
Figure 79
Connection Diagram for ISI J-20 Calibration