Keysight N5990A MIPI M-PHY Receiver Test User Guide
125
Calibrations
4
SJ Calibration
This calibration is available for all hardware configurations.
Figure 71
SJ Calibration
Purpose
It calibrates the low frequency SJ that is injected internally by the BERT.
This calibration is run once at each selected HS data rate (see
For spec 4.1, this calibration is only mandatory for B-series (for A-series,
the procedure is shown in the test tree as informative).