Keysight N5990A MIPI M-PHY Receiver Test User Guide
133
Calibrations
4
Figure 75
Example HTML Viewer for Eye Opening Height
Table 16
Eye Opening Calibration with Jitter data table
Parameter Name
Parameter Description
Result
"Pass"/"Fail" - If the set-up is wrong, the calibration procedure fails due to the non-monotonic values
measured.
SJ Frequency
It is the set value for sinusoidal jitter frequency at each step.
SJ Amplitude
It is the set value for sinusoidal jitter amplitude.
Eye Width
It is the measured value of eye width according to the set SJ.
Voltage Amplitude
It is the set value for voltage amplitude.
Eye Height
It is the measured value of eye height according to the set voltage.
Maximum V_diff_ac_tx
It is the maximum measured value of the differential voltage amplitude.
Содержание N5990A
Страница 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Страница 17: ...Keysight N5990A MIPI M PHY Receiver Test User Guide 1 Introduction Document History 18 Overview 19...
Страница 62: ......
Страница 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Страница 214: ......
Страница 234: ......
Страница 268: ......
Страница 298: ......