99
Calibrations
4
ISI (w/Switch) Calibration
This procedure is available for all hardware configurations.
Figure 54
ISI Calibration
Purpose
The receiver test setup must include an ISI Compliance Channel, capable
of introducing a channel loss as defined in the M-PHY specification. This
procedure calibrates the ISI of the trace used for that purpose.
This calibration should run once at each data rate (see
).