Keysight N5990A MIPI M-PHY Receiver Test User Guide
251
PWM Tests
7
Parameter Name
Parameter Description
Result
"Pass"/"Fail" - If the value is "Fail", the termination enable time is not within the spec limit.
PWM Data Rate
It is the PWM data rate value in MBit/s.
Termination Time Min
It is the shortest/minimum termination enable time in "ns" (nanosecond).
Termination UI Min
It is the shortest/minimum termination enable time in "UI".
Min Spec
It is the minimum value of the termination enable time, where the DUT must work according to the
specification.
Termination Time Max
It is the maximum termination enable time in "ns".
Termination UI Max
It is the maximum termination enable time in "UI".
Max Spec
It is the maximum value of the termination enable time, where the DUT must work according to the
specification.