134
Keysight N5990A MIPI M-PHY Receiver Test User Guide
4
Calibrations
Accumulated Voltage Calibration Terminated
It is available for all hardware configurations.
Figure 76
Accumulated Voltage Calibration terminated / Into Open
Purpose
It calibrates the accumulated differential voltage, V
DIF-ACC-RX
, with only
the Reference ISI trace in the signal path for the Test 2.1.2-HS-RX
Accumulated Differential Input Voltage Tolerance Test usage.
This calibration should be run once at each HS selected data rate (see
When the
DSGA is used, two calibrations are required for each data rate:
one for Data0 and another one for the other channels. The calibration for
Data0 includes the switch. For other channels, Data1 is calibrated without
the switch.