
Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-52
2461-901-01 A/November 2015
digio.line[N].reset()
This function resets digital I/O line values to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
digio.line[
N
].reset()
N
The digital I/O line (1 to 6)
Details
This function resets the following attributes to their default values:
•
digio.line[
N
].mode
•
trigger.digin[
N
].edge
•
trigger.digout[
N
].logic
•
trigger.digout[
N
].pulsewidth
•
trigger.digout[
N
].stimulus
It also clears
trigger.digin[
N
].overrun
.
Example
-- Set the digital I/O trigger line 3 for a falling edge
digio.line[3].mode = digio.MODE_TRIGGER_OUT
trigger.digout[3].logic = trigger.LOGIC_NEGATIVE
-- Set the digital I/O trigger line 3 to have a pulsewidth of 50 microseconds.
trigger.digout[3].pulsewidth = 50e-6
-- Use digital I/O line 5 to trigger the event on line 3.
trigger.digout[3].stimulus = trigger.EVENT_DIGIO5
-- Print configuration (before reset).
print(digio.line[3].mode, trigger.digout[3].pulsewidth, trigger.digout[3].stimulus)
-- Reset the line back to factory default values.
digio.line[3].reset()
-- Print configuration (after reset).
print(digio.line[3].mode, trigger.digout[3].pulsewidth, trigger.digout[3].stimulus)
Output before reset:
digio.MODE_TRIGGER_OUT
5e-05 trigger.EVENT_DIGIO5
Output after reset:
digio.MODE_TRIGGER_IN
1e-05 trigger.EVENT_NONE
Also see
(on page 8-50)
Digital I/O port configuration
(on page 3-87)
(on page 8-243)
(on page 8-245)
(on page 8-247)