Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-18
2461-901-01 A/November 2015
:ACAL:LASTrun:TIME?
This command returns the date and time when autocalibration was last run.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:ACAL:LASTrun:TIME?
Details
The date and time is returned in the format:
MM/DD/YYYY HH:MM:SS.NNNNNNNNN
Where:
•
MM/DD/YYYY
is the month, date, and year
•
HH:MM:SS.NNNNNNNNN
is the hour, minute, second, and fractional second
Example
ACAL:LAST:TIME?
Returns the date and time when auto calibration
was last run.
Example output:
08/11/2014 16:30:26.745369595
Also see
(on page 6-18)
:ACAL:RUN
This command immediately runs autocalibration and stores the constants.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Nonvolatile memory
Not applicable
Usage
:ACAL:RUN
Details
During autocalibration, all necessary autocalibration steps are completed, calibration constants are
updated, and autocalibration last run information is updated.
When an autocalibration command is received, the instrument runs the autocalibration to completion
before executing the next command.
Example
ACAL:RUN
Auto calibration starts running.
Also see
(on page 3-51)