Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-10
2461-901-01 A/November 2015
Example
acal.run()
Autocalibration starts running.
Also see
(on page 3-51)
beeper.beep()
This function generates an audible tone.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
beeper.beep(
duration
,
frequency
)
duration
The amount of time to play the tone (0.001 to 100 s)
frequency
The frequency of the beep (20 to 8000 Hz)
Details
You can use the beeper of the instrument to provide an audible signal at a specific frequency and
time duration. For example, you can use the beeper to signal the end of a lengthy sweep.
Using this function from a remote interface does not affect audible errors or key click settings that
were made from the Model 2461 front panel.
Example
beeper.beep(2, 2400)
Generates a 2 s, 2400 Hz tone.
Also see
None
buffer.clearstats()
This function clears the statistical information associated with the specified buffer.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
buffer.clearstats()
buffer.clearstats(
bufferVar
)
bufferVar
The name of the reading buffer, which may be a default buffer (
defbuffer1
or
defbuffer2
) or a user-defined buffer; defaults to
defbuffer1
if not specified
Details
This command clears the statistics without clearing the readings.