Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2461-901-01 A/November 2015
6-169
Example
TRACe:MAKE "MyBuffer", 100
SENSe:COUNt 5
MEASure:CURRent:DC? "MyBuffer"
TRACe:DATA? 1,5, "MyBuffer", READ, REL, SOUR
TRACe:SAVE "/usb1/myData.csv", "MyBuffer"
TRACe:SAVE "/usb1/myDataRel.csv", "MyBuffer", REL
Create a buffer called
MyBuffer
with a
maximum size of 100.
Make five readings for each
measurement request and return the
data.
Make the measurements.
Read the reading, relative timestamp,
and source value for each point from 1
to 5.
Output:
-0.000000,0.000000,
0.000000,-0.000000,
0.301759,0.000000,
-0.000000,0.579068,
0.000000,-0.000000,
0.884302,0.000000,
-0.000000,1.157444,
0.000000
Save all reading and default time
information from a buffer named
MyBuffer
to a file named
myData.csv
on the USB flash drive.
Save all readings and relative
timestamps from
MyBuffer
to a file
named
myDataRel.csv
on the USB
flash drive.
Also see
(on page 3-2)
(on page 3-23)
(on page 6-165)
(on page 6-170)