To create a sweep:
1. Set the source function using
smu.source.func
.
2. Set the source range using
smu.source.range
.
3. Set any other source settings that apply to your sweep. You must set source settings before the
sweep function is called.
4. If you are using
smu.source.sweeplist()
, set up the source configuration list for your sweep.
5. Set the parameters for the sweep command.
6. Set the measurement function using
smu.measure.func
.
7. Set the measurement range using
smu.measure.range
.
8. Make any other settings appropriate to your sweep.
9. Send
trigger.model.initiate()
to start the sweep.
To save your settings, save them to a configuration script using the
createconfigscript()
command.
For example sweeps, see
(on page 3-61).
For detail on the commands and options listed above, see the following command descriptions:
•
(on page 8-210)
•
(on page 8-212)
•
(on page 8-214)
•
(on page 8-216)
•
(on page 8-184)
•
(on page 8-207)
•
smu.source.configlist.create()
(on page 8-177)
•
(on page 8-182)
•
(on page 8-150)
•
(on page 8-166)
Aborting a sweep
Sweeps can be stopped for the following reasons:
•
The limit set by the abort on limit setting was exceeded
•
The trigger model is aborted
You can stop the sweep while it is in progress. When you stop the sweep, all sweep commands in the
trigger model are terminated.