Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2461-901-01 A/November 2015
6-17
:ACAL:LASTrun:TEMPerature:INTernal?
This command returns the internal temperature of the instrument when autocalibration was run.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Nonvolatile memory
Not applicable
Usage
:ACAL:LASTrun:TEMPerature:INTernal?
Details
The temperature is displayed in Celsius (°C).
Example
ACAL:LAST:TEMP:INT?
Returns the internal temperature of the
instrument when autocalibration was last run.
Example output:
63.167084
Also see
(on page 6-18)
(on page 3-51)
:ACAL:LASTrun:TEMPerature:DIFFerence?
This command returns the difference between the internal temperature and the temperature when autocalibration
was last run.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:ACAL:LASTrun:TEMPerature:DIFFerence?
Details
The temperature is displayed in Celsius (°C).
Example
ACAL:LAST:TEMP:DIFF?
Returns the difference between the temperature
of the instrument when autocalibration was last
run and the present internal temperature.
Example output:
4.5678
Also see
:ACAL:LASTrun:TEMPerature:INTernal?
(on page 6-17)
(on page 6-18)
(on page 3-51)