Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-139
Example
print(smu.measure.configlist.query("testMeasList", 2, "\n"))
Returns the TSP commands and parameter settings that represent the settings in configuration index 2.
Partial example output:
smu.measure.func = smu.FUNC_DC_CURRENT
smu.measure.range = 1.000000e-06
smu.measure.autorange = smu.ON
smu.measure.autorangelow = 1.000000e-06
smu.measure.autozero.enable = smu.ON
smu.measure.displaydigits = smu.DIGITS_5_5
smu.measure.filter.enable = smu.OFF
smu.measure.filter.count = 10
smu.measure.filter.type = smu.FILTER_REPEAT_AVG
smu.measure.limit[1].autoclear = smu.ON
smu.measure.limit[1].enable = smu.OFF
smu.measure.limit[1].audible = smu.AUDIBLE_NONE
smu.measure.limit[1].high.value = 1.00
smu.measure.limit[1].low.value = -1.00
smu.measure.limit[2].autoclear = smu.ON
smu.measure.limit[2].audible = smu.AUDIBLE_NONE
smu.measure.limit[2].enable = smu.OFF
smu.measure.limit[2].high.value = 1.00
smu.measure.limit[2].low.value = -1.00
smu.measure.math.enable = smu.OFF
smu.measure.math.format = smu.MATH_PERCENT
smu.measure.math.mxb.bfactor = 0.00
smu.measure.math.mxb.mfactor = 1.00
smu.measure.math.percent = 1.00
smu.measure.userdelay[1] = 0.00
smu.measure.userdelay[2] = 0.00
smu.measure.userdelay[3] = 0.00
smu.measure.userdelay[4] = 0.00
smu.measure.userdelay[5] = 0.00
smu.measure.nplc = 1.00
smu.measure.offsetcompensation = smu.OFF
smu.measure.autorangehigh = is not used
smu.measure.sense = smu.SENSE_2WIRE
smu.measure.unit = smu.UNIT_AMP
smu.measure.rel.enable = smu.OFF
smu.measure.rel.level = 0.00
smu.digitize.func = smu.FUNC_NONE
smu.digitize.range = is not used
smu.digitize.displaydigits = is not used
Also see
(on page 3-30)
smu.measure.configlist.create()
(on page 8-136)