Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-238
2461-901-01 A/November 2015
trigger.blender[N].reset()
This function resets some of the trigger blender settings to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.blender[
N
].reset()
N
The trigger event blender (1 or 2)
Details
The
trigger.blender[
N
].reset()
function resets the following attributes to their factory
defaults:
•
trigger.blender[
N
].orenable
•
trigger.blender[
N
].stimulus[
M
]
It also clears
trigger.blender[
N
].overrun
.
Example
trigger.blender[1].reset()
Resets the trigger blender 1
settings to factory defaults.
Also see
(on page 8-236)
(on page 8-237)
trigger.blender[N].stimulus[M]
trigger.blender[N].stimulus[M]
This attribute specifies the events that trigger the blender.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Trigger blender N reset
Configuration script
trigger.EVENT_NONE
Usage
event
= trigger.blender[
N
].stimulus[
M
]
trigger.blender[
N
].stimulus[
M
] =
event
event
The event that triggers the blender action; see
Details
N
An integer that represents the trigger event blender (1 or 2)
M
An integer representing the stimulus index (1 to 4)
Details
There are four stimulus inputs that can each select a different event.
Use zero to disable the blender input.