Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2461-901-01 A/November 2015
3-59
5. Set the parameters for the sweep command.
6. Set the measurement function using
[:SENSe[1]]:FUNCtion
.
7. Set the measurement range using
[:SENSe[1]]:<function>:RANGe[:UPPer]
.
8. Make any other settings appropriate to your sweep.
9. Send
:INITiate
to start the sweep.
To save your settings, save them to a user-saved setup using the *
SAV
command.
For example sweeps, see
(on page 3-61).
For detail on the commands and options listed above, see the following command descriptions:
•
(on page 6-81)
•
[:SENSe[1]]:<function>:RANGe[:UPPer]
(on page 6-65)
•
(on page 6-95)
•
(on page 6-97)
•
:SOURce[1]:CONFiguration:LIST:CREate
(on page 6-83)
•
:SOURce[1]:CONFiguration:LIST:STORe
(on page 6-88)
•
:SOURce[1]:SWEep:<function>:LINear
(on page 6-124)
•
:SOURce[1]:SWEep:<function>:LINear:STEP
(on page 6-126)
•
:SOURce[1]:SWEep:<function>:LIST
(on page 6-128)
•
:SOURce[1]:SWEep:<function>:LOG
(on page 6-130)
Setting up a sweep using TSP commands
To set up a sweep using TSP commands, you send one of the following commands:
•
smu.source.sweeplinear()
: Sets up a linear sweep for a fixed number of measurement
points.
•
smu.source.sweeplinearstep()
: Sets up a linear source sweep configuration list and
trigger model with a fixed number of steps.
•
smu.source.sweeplist()
: Sets up a sweep based on a configuration list, which allows you to
customize the sweep.
•
smu.source.sweeplog()
: Sets up a logarithmic sweep for a set number of measurement
points.