Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-183
smu.source.delay
This attribute contains the source delay.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Source configuration list
Function change
Configuration script
Source configuration list
Not applicable
Usage
sDelay
= smu.source.delay
smu.source.delay =
sDelay
sDelay
The length of the delay (0 to 10 ks)
Details
This command sets a delay for the selected source function. This delay is in addition to normal
settling times.
After the programmed source is turned on, this delay allows the source level to settle before a
measurement is made.
If you set a specific source delay (
smu.source.delay
), source autodelay is turned off.
When source autodelay is turned on, the manual source delay setting is overwritten with the
autodelay setting.
When either a source delay or autodelay is set, the delay is applied to the first source output and then
only when the magnitude of the source changes.
Example
smu.source.func = smu.FUNC_DC_VOLTAGE
smu.source.delay = 3
Set the function to voltage. Set a 3 s delay
after the source is turned on before a
measurement is taken.
Also see
(on page 8-176)
(on page 2-130)