Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-157
Example output that shows all readings are within limit values (all readings between 3 V and 5 V):
limit 1 results = smu.FAIL_NONE
limit 2 results = smu.FAIL_NONE
Example output showing at least one reading failed limit 1 high values (a 6 V reading would cause this
condition or a reading greater than 5 V but less than 7 V):
limit 1 results = smu.FAIL_HIGH
limit 2 results = smu.FAIL_NONE
Example output showing at least one reading failed limit 1 and 2 low values (a 0.5 V reading would cause this
condition or a reading less than 1 V):
limit 1 results = smu.FAIL_LOW
limit 2 results = smu.FAIL_LOW
Also see
(on page 8-36)
(on page 3-134)
(on page 8-154)
smu.measure.limit[Y].high.value
This attribute specifies the upper limit for a limit test.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
1
Usage
highLimit
= smu.measure.limit[
Y
].high.value
smu.measure.limit[
Y
].high.value =
highLimit
highLimit
The value of the high limit (
−
9.11 to +9.11)
Y
Limit number:
1
or
2
Details
This command sets the high limit for the limit
Y
test for the selected measurement function. When limit
Y
testing is enabled, the instrument generates a fail indication when the measurement value is more
than this value.
Example
(on page 8-155).
Also see
(on page 8-154)