Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2461-901-01 A/November 2015
6-109
:SOURce[1]:PULSe:LIST:<function>:POINts?
This command returns the number of configuration indexes in the source pulse list for the selected source
function.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:SOURce[1]:PULSe:LIST:CURRent:POINts?
:SOURce[1]:PULSe:LIST:VOLTage:POINts?
Details
This command returns the length of the
CurrPulseCustomSweepList
if the function is current or
VoltPulseCustomSweepList
if the function is voltage. The response message indicates the
number of source values in the list.
Example
*RST
SENS:FUNC "CURR"
SENS:CURR:RANG:AUTO ON
SENS:CURR:RSEN OFF
SOUR:FUNC VOLT
SOUR:VOLT:RANG 20
SOUR:PULS:VOLT:ILIM 1
SOUR:PULS:LIST:VOLT 1, 5, 1, 5, 1, 5
SOUR:PULS:SWE:VOLT:LIST 200e-3, ON, "defbuffer1", 1, 3, 0, 200e-3, ON
INIT
*WAI
TRAC:DATA? 1, 6, "defbuffer1", SOUR, READ
SOUR:PULS:LIST:VOLT:POIN?
This example will source 1 V, 5 V, 1 V, 5 V, 1 V, 5 V and measure the resulting current at each voltage point.
The time duration of each voltage point is 200 ms.
Check the number of points in the list.
Output from the
SOUR:PULS:LIST:VOLT:POIN?
query:
6
Also see
(on page 3-64)
(on page 6-101)
:SOURce[1]:LIST:<function>:APPend
(on page 6-102)
:SOURce[1]:LIST:<function>:POINts?
(on page 6-103)
:SOURce[1]:PULSe:LIST:<function>
(on page 6-107)
:SOURce[1]:PULSe:LIST:<function>:APPend
(on page 6-108)
(on page 3-107)