Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-196
2461-901-01 A/November 2015
Example
reset()
smu.source.func = smu.FUNC_DC_VOLTAGE
smu.source.range = 20
smu.measure.func = smu.FUNC_DC_CURRENT
smu.measure.range = 100e-6
smu.measure.nplc = 0.01
smu.measure.autozero.enable = smu.OFF
smu.source.pulsesweeplinear("VoltLinSweep", 0, 0, 10, 20, 2e-3)
trigger.model.initiate()
Reset the instrument to its default settings.
Set the source function to voltage and the source range to 20 V.
Set up a linear pulse sweep that sweeps from 0 to 10 V in 20 steps with a pulse width of 2 ms and a sweep
count of 1. Name the configuration list that is created for this sweep
VoltLinSweep
.
Set the measure function to current and the current range to 100
µ
A.
Start the pulse sweep.
Also see
(on page 8-200)
(on page 3-53)
(on page 8-259)