Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-78
2461-901-01 A/November 2015
Details
This command returns the size (number of configuration indexes) of a measure configuration list. The
size of the list is equal to the number of configuration indexes in a configuration list.
Example
:SENSe:CONF:LIST:SIZE? "MyMeasList"
Returns the number of configuration indexes
in a measure configuration list named
MyMeasList
.
Example output:
3
Also see
(on page 3-30)
[:SENSe[1]]:CONFiguration:LIST:CREate
(on page 6-74)
[:SENSe[1]]:CONFiguration:LIST:STORe
This command stores the active measure or digitize settings into the named configuration list.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Saved settings
Not applicable
Usage
[:SENSe[1]]:CONFiguration:LIST:STORe "<name>"
[:SENSe[1]]:CONFiguration:LIST:STORe "<name>", <index>
<name>
A string that represents the name of a measure configuration list
<index>
A number that defines a specific configuration index in the configuration list
Details
Use this command to store the active settings to a configuration index in a configuration list. If you do
not include the
<index>
parameter, the configuration index is appended to the end of the list.
Refer to
Instrument settings stored in a measure configuration list
(on page 3-33) for a complete list of
measure settings that the instrument stores.
Example
:SENSe:CONF:LIST:STOR "MyConfigList"
Stores the active settings of the instrument to the
end of the configuration list named
MyConfigList
.
:SENSe:CONF:LIST:STOR "MyConfigList", 5
Stores the active settings of the instrument to the
configuration list named
MyConfigList
in
configuration index 5.
Also see
(on page 3-30)
[:SENSe[1]]:CONFiguration:LIST:CREate
(on page 6-74)