
Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2461-901-01 A/November 2015
3-43
The following table provides information about specific settings for each configuration index.
Settings
Index 1
Index 2
Index 3
Index 4
Function
Voltage
Voltage
Voltage
Voltage
AutoRange
Off
Off
Off
Off
Delay
0.001
0.001
0.001
0.001
AutoDelay
Off
Off
Off
Off
Source Level
2.0 V
3.0 V
4.0 V
5.0 V
ProtectLevel
PROTECT_40
PROTECT_40
PROTECT_40
PROTECT_40
ProtectTripped
Not tripped
Not tripped
Not tripped
Not tripped
OffMode
OFFMODE_
NORMAL
OFFMODE_
NORMAL
OFFMODE_
NORMAL
OFFMODE_
NORMAL
Range
20.0 V
20.0 V
20.0 V
20.0 V
LimitLevel
10.6 nA
10.6 nA
10.6 nA
10.6 nA
LimitTripped
Not tripped
Not tripped
Not tripped
Not tripped
HighC
Off
Off
Off
Off
Readback
On
On
On
On
UserDelay1-5
None
None
None
None
The figures in this section include a graphic representation of
MySourceList
and the four
configuration indexes. For simplicity, the figures will only show two of the source settings listed in the
table (source range and source level).
This example creates a source configuration list; however, you can use the same process to create a
measure configuration list.
Creating a configuration list and giving it a name
Use one of the following methods to create the configuration list used in this example and give it a
name.
Using SCPI commands:
:SOURce:CONFiguration:LIST:CREate "MySourceList"
Using TSP commands:
smu.source.configlist.create("MySourceList")