Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-42
2461-901-01 A/November 2015
:DISPlay:READing:FORMat
This command determines the format that is used to display measurement readings on the front-panel display of
the instrument.
Type
Affected by
Where saved
Default value
Command and query
Not applicable
Nonvolatile memory
PREF
Usage
:DISPlay:READing:FORMat <format>
:DISPlay:READing:FORMat?
<format>
Use exponent format:
EXPonent
Add a prefix to the units symbol, such as k, m, or µ:
PREFix
Details
This setting persists through
*RST
and power cycles.
When Prefix is selected, prefixes are added to the units symbol, such as k (kilo) or m (milli). When
Exponent is selected, exponents are used instead of prefixes. When the prefix option is selected, very
large or very small numbers may be displayed with exponents.
Example
DISP:READ:FORM EXP
Change front-panel display to show readings in
exponential format.
Also see
(on page 2-46)
:DISPlay:SCReen
This command changes which front-panel screen is displayed.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:DISPlay:SCReen <screenName>
<screenName>
The screen to display:
•
Home screen:
HOME
•
Home screen with large readings:
HOME_LARGe_reading
•
Reading table:
READing_table
•
Graph screen (opens last selected tab):
GRAPh
•
Histogram screen:
HISTogram
•
GRAPH swipe screen:
SWIPE_GRAPh
•
SETTINGS swipe screen:
SWIPE_SETTings
•
SOURCE swipe screen:
SOURce
•
STATISTICS swipe screen:
SWIPE_STATistics
•
USER swipe screen:
SWIPE_USER