Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2461-901-01 A/November 2015
6-21
Details
This attribute specifies the offset (b) for an mx + b operation.
The mx + b math operation lets you manipulate normal display readings (x) mathematically based on
the calculation:
y = mx + b
Where:
•
y
is the displayed result
•
m
is a user-defined constant for the scale factor
•
x
is the measurement reading (if you are using a relative offset, this is the measurement with relative
offset applied)
•
b
is the user-defined constant for the offset factor
If you send this command without the
<function>
parameter, it will set the scale factor for all
measure functions. It will not change the setting for a digitize function.
Example
:CALC:VOLT:MATH:FORM MXB
:CALC:VOLT:MATH:MMF 0.80
:CALC:VOLT:MATH:MBF 50
:CALC:VOLT:MATH:STAT ON
Set the math function for voltage measurements
to mx+b.
Set the scale factor for voltage measurements to
0.80.
Set the offset factor to 50.
Enable the math function.
Also see
Calculations that you can apply to measurements
(on page 3-82)
:CALCulate[1]:<function>:MATH:FORMat
(on page 6-19)
:CALCulate[1]:<function>:MATH:MMFactor
(on page 6-22)
:CALCulate[1]:<function>:MATH:STATe
(on page 6-24)