Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-22
2461-901-01 A/November 2015
:CALCulate[1]:<function>:MATH:MMFactor
This command specifies the scale factor, m, for the y = mx + b math operation.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
1
Usage
:CALCulate[1]:<function>:MATH:MMFactor <n>
:CALCulate[1]:<function>:MATH:MMFactor DEFault
:CALCulate[1]:<function>:MATH:MMFactor MINimum
:CALCulate[1]:<function>:MATH:MMFactor MAXimum
:CALCulate[1]:<function>:MATH:MMFactor?
:CALCulate[1]:<function>:MATH:MMFactor? DEFault
:CALCulate[1]:<function>:MATH:MMFactor? MINimum
:CALCulate[1]:<function>:MATH:MMFactor? MAXimum
<function>
The measure function:
•
Current:
CURRent[:DC]
•
Resistance:
RESistance
•
Voltage:
VOLTage[:DC]
The digitize function:
•
Digitize current:
DIGitize:CURRent
•
Digitize voltage:
DIGitize:VOLTage
<n>
The scale factor; the valid range is
−
1e12 to
+
1e12
Details
This command sets the scale factor (m) for an mx + b operation for the selected measurement
function.
The mx + b math operation lets you manipulate normal display readings (x) mathematically according
to the following calculation:
y = mx + b
Where:
•
y
is the displayed result
•
m
is a user-defined constant for the scale factor
•
x
is the measurement reading (if you are using a relative offset, this is the measurement with relative
offset applied)
•
b
is the user-defined constant for the offset factor
If you send this command without the
<function>
parameter, it will set the scale factor for all
measure functions. It will not change the setting for a digitize function.