Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2461-901-01 A/November 2015
6-63
Example
RES:RANG:AUTO ON
Set the range to be selected automatically
for resistance measurements.
Also see
[:SENSe[1]]:<function>:RANGe[:UPPer]
(on page 6-65)
[:SENSe[1]]:<function>:RANGe:AUTO:LLIMit
(on page 6-63)
[:SENSe[1]]:<function>:RANGe:AUTO:LLIMit
This command selects the lower limit for measurements of the selected function when the range is selected
automatically.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
Current: 1 µA
Voltage: 200 mV
Resistance: 2
Ω
Usage
[:SENSe[1]]:<function>:RANGe:AUTO:LLIMit <n>
[:SENSe[1]]:<function>:RANGe:AUTO:LLIMit?
[:SENSe[1]]:<function>:RANGe:AUTO:LLIMit? DEFault
[:SENSe[1]]:<function>:RANGe:AUTO:LLIMit? MINimum
[:SENSe[1]]:<function>:RANGe:AUTO:LLIMit? MAXimum
<function>
The measure function:
•
Current:
CURRent[:DC]
•
Resistance:
RESistance
•
Voltage:
VOLTage[:DC]
<n>
The lower limit:
•
Current: 1 µA to 5 A
•
Voltage: 200 mV to 20 V
•
Resistance: 2
Ω
to 20 M
Ω