Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-136
2461-901-01 A/November 2015
Example
:STATus:OPERation:MAP 0, 4916, 4917
When event 4916 (the buffer is 0 % filled)
occurs, bit 0 is set in the condition register and
the event register of the Operation Event
Register. When event 4917 (buffer is 100 %
filled) occurs, bit 0 in the condition register is
cleared.
Also see
Programmable status register sets
(on page C-5)
:STATus:PRESet
This command resets all bits in the status model.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:STATus:PRESet
Details
This function clears the event registers and the enable registers for operation and questionable. It will
not clear the Service Request Enable Register (
*SRE
) to Standard Request Enable Register (
*ESE
).
Preset does not affect the event queue.
The Standard Event Status Register is not affected by this command.
Example
STAT:PRES
Resets the registers.
Also see
(on page C-1)
:STATus:QUEStionable:CONDition?
This command reads the Questionable Condition Register of the status model.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:STATus:QUEStionable:CONDition?
Details
This command reads the contents of the Questionable Condition Register, which is one of the
Questionable Event Registers.
For detail on interpreting the value of a register, see
(on page C-16).