Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2461-901-01 A/November 2015
6-69
[:SENSe[1]]:<function>:RELative:STATe
This command enables or disables the application of a relative offset value to the measurement.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0 (OFF)
Usage
[:SENSe[1]]:<function>:RELative:STATe <state>
[:SENSe[1]]:<function>:RELative:STATe?
<function>
The measure function:
•
Current:
CURRent[:DC]
•
Resistance:
RESistance
•
Voltage:
VOLTage[:DC]
The digitize function:
•
Digitize current:
DIGitize:CURRent
•
Digitize voltage:
DIGitize:VOLTage
<state>
Disable the relative offset:
OFF
or
0
Enable the relative offset:
ON
or
1
Details
When relative measurements are enabled, all subsequent measured readings are offset by the
relative offset value. You can enter a relative offset value or have the instrument acquire a relative
offset value.
Each returned measured relative reading is the result of the following calculation:
Displayed reading = Actual measured reading - Relative offset value
Example
:SENS:FUNC "VOLT"
:SENS:VOLT:REL 5
:SENSe:VOLT:REL:STATe ON
Set the measurement function to volts with a
relative offset of 5 V and enable the relative
offset function.
Also see
(on page 3-79)
[:SENSe[1]]:<function>:RELative
(on page 6-67)
[:SENSe[1]]:<function>:RELative:ACQuire
(on page 6-68)