Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-172
2461-901-01 A/November 2015
Details
This command returns the average reading calculated from all of the readings in the specified reading
buffer.
When the reading buffer is configured to fill continuously and overwrite old data with new data, the
buffer statistics include the data that was overwritten. To get statistics that do not include data that
has been overwritten, define a large buffer size that will accommodate the number of readings you
will make.
Example
TRACe:STAT:AVERage?
Returns the average reading for the readings in the default
buffer
defbuffer1
.
TRACe:STAT:AVERage? "testData"
Returns the average reading for the readings in the user-
defined buffer
testData
.
Also see
(on page 3-2)
(on page 3-23)
(on page 6-165)
(on page 6-172)
(on page 6-173)
(on page 6-174)
(on page 6-175)
(on page 6-175)
:TRACe:STATistics:CLEar
This command clears the statistical information associated with the specified buffer.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:TRACe:STATistics:CLEar
:TRACe:STATistics:CLEar "<bufferName>"
<bufferName>
The name of the reading buffer, which may be a default buffer
(
defbuffer1
or
defbuffer2
) or a user-defined buffer; if no buffer is
defined, clears the statistics from
defbuffer1
Details
This command clears the statistics without clearing the readings.