Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-269
Trigger events
Event description
Event constant
Trigger event blender
N
(1 to 2), which combines trigger
events
trigger.EVENT_BLENDER
N
Trigger timer
N
(1 to 4) expired
trigger.EVENT_TIMER
N
Source limit condition occurs
trigger.EVENT_SOURCE_LIMIT
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel
MENU
key and under Trigger, selecting
Configure
. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
trigger.model.getblocklist()
command to view the trigger model blocks in a list format.
Example
reset()
-- Set up measure function
smu.measure.func = smu.FUNC_DC_CURRENT
-- Initiate readings
trigger.model.load("LoopUntilEvent", trigger.EVENT_DISPLAY, 50)
trigger.model.initiate()
Reset the instrument.
Set the instrument to measure current.
Load the LoopUntilEvent trigger model to make measurements until the front panel trigger key is pressed,
then continue to make measurements equal to 50 % of the reading buffer size.
Start the trigger model.
Also see