Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-160
2461-901-01 A/November 2015
Example
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.math.format = smu.MATH_RECIPROCAL
smu.measure.math.enable = smu.ON
Enables the reciprocal math operation on
voltage measurements.
Also see
Calculations that you can apply to measurements
(on page 3-82)
(on page 8-158)
(on page 8-160)
(on page 8-162)
(on page 8-163)
smu.measure.math.mxb.bfactor
This attribute specifies the offset, b, for the y = mx + b operation.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
0
Usage
value
= smu.measure.math.mxb.bfactor
smu.measure.math.mxb.bfactor =
value
value
The offset for the y = mx + b operation; the valid range is
−
1e12 to
+
1e12
Details
This attribute specifies the offset (b) for an mx + b operation.
The mx + b math operation lets you manipulate normal display readings (x) mathematically based on
the calculation:
y = mx + b
Where:
•
y
is the displayed result
•
m
is a user-defined constant for the scale factor
•
x
is the measurement reading (if you are using a relative offset, this is the measurement with relative
offset applied)
•
b
is the user-defined constant for the offset factor