Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-214
2461-901-01 A/November 2015
smu.source.sweeplist()
This function sets up a sweep based on a configuration list, which allows you to customize the sweep.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
smu.source.sweeplist(
configListName
)
smu.source.sweeplist(
configListName
,
index
)
smu.source.sweeplist(
configListName
,
index
,
sDelay
)
smu.source.sweeplist(
configListName
,
index
,
sDelay
,
count
)
smu.source.sweeplist(
configListName
,
index
,
sDelay
,
count
,
failAbort
)
smu.source.sweeplist(
configListName
,
index
,
sDelay
,
count
,
failAbort
,
bufferName
)
configListName
The name of the source configuration list that the sweep uses; this must be defined
before sending this command
index
The index in the configuration list where the sweep starts; default is
1
sDelay
The delay between measurement points; default is
0
for no delay or you can set a
specific delay value from 50
µ
s to 10 ks
count
The number of times to run the sweep; default is
1
:
•
Infinite loop:
smu.INFINITE
•
Finite loop:
1
to
268,435,455
failAbort
Complete the sweep if the source limit is exceeded:
smu.OFF
Abort the sweep if the source limit is exceeded:
smu.ON
(default)
bufferName
Indicates the reading buffer to use; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1
is used
Details
This command allows you to set up a custom sweep, using a configuration list to specify the source
levels.
When you specify a delay, a delay block is added to the sweep trigger model. This delay is added to
any source delay you may have set. For example, if you set 10 ms for the source delay and 25 ms for
the delay in the for the log sweep command, the actual delay is 35 ms.
To run the sweep, initiate the trigger model.