Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-247
trigger.digout[N].stimulus
This attribute selects the event that causes a trigger to be asserted on the digital output line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Digital I/O trigger
N
reset
Configuration script
trigger.EVENT_NONE
Usage
event
= trigger.digout[
N
].stimulus
trigger.digout[
N
].stimulus =
event
event
The event to use as a stimulus; see
Details
N
Digital I/O trigger line (1 to 6)
Details
The digital trigger pulsewidth command determines how long the trigger is asserted.
The trigger stimulus for a digital I/O line can be set to one of the trigger events that are described in
the following table.
Trigger events
Event description
Event constant
No trigger event
trigger.EVENT_NONE
Front-panel TRIGGER key press
trigger.EVENT_DISPLAY
Notify trigger block
N
(1 to 8) generates a trigger event when
the trigger model executes it
trigger.EVENT_NOTIFY
N
A command interface trigger:
•
Any remote interface: *
TRG
•
GPIB only: GET bus command
•
VXI-11: VXI-11 command
device_trigger
trigger.EVENT_COMMAND
Line edge (either rising, falling, or either based on the
configuration of the line) detected on digital input line
N
(1 to 6)
trigger.EVENT_DIGIO
N
Line edge detected on TSP-Link synchronization line
N
(1 to 3)
trigger.EVENT_TSPLINK
N
Appropriate LXI trigger packet is received on LAN trigger
object
N
(1 to 8)
trigger.EVENT_LAN
N