![JDS Uniphase CellAdvisor JD780A Series Скачать руководство пользователя страница 77](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616077.webp)
JD780A Series Analyzers
Using Spectrum Analyzer
77
d.
Toggle the
Link
soft key to select the sweep direction and select
Fwd
or
Rev
as
needed.
e.
Press the
Lowest Channel
soft key.
f.
Enter a value by using the numeric keys.
g.
Press the
Enter
soft key or
ENTER
hard key to complete the entry.
The instrument automatically displays the corresponding center frequency value for the
selected channel number.
h.
Press the
Highest Channel
soft key.
i.
Enter a value by using the numeric keys.
j.
Press the
Enter
soft key or
ENTER
hard key to complete the entry.
The instrument automatically displays the corresponding center frequency value for the
selected channel number.
3.
To set up ACP setting, press the
ACP Settings
soft key and then complete the following steps:
a.
Press the
Main Channel BW
soft key.
b.
Enter a value by using the numeric keys. You can also use the rotary knob.
c.
Select the unit:
MHz
,
kHz
, or
Hz
.
d.
Press the
Span
soft key to set the frequency range over which the instrument will sweep.
e.
Enter a span frequency value by using the numeric keys. You can also use the rotary knob.
f.
Select the unit:
GHz
,
MHz
,
kHz
, or
Hz
.
g.
Press the
Average
soft key to set the number of measurement to be averaged:
h.
Enter a value between 1 and 100.
i.
Press the
Enter
soft key.
4.
To configure an offset, up to five, complete the following steps:
a.
Press the
Offset/Limit
soft key. The Offset/Limit screen menu bar appears.
b.
Press the
Offset
soft key, and then select the number for the active offset to configure.
c.
Toggle the
Offset
soft key and select
On
to display or
Off
to hide the active offset.
d.
Press the
Offset Frequency
soft key to set the offset frequency.
e.
Enter a value by using the numeric keys. You can also use the rotary knob.
f.
Select the unit:
MHz
,
kHz
, or
Hz
.
g.
Press the
Integration Bandwidth
soft key.
h.
Enter a value by using the numeric keys. You can also use the rotary knob.
i.
Select the unit:
MHz
,
kHz
, or
Hz
.
5.
To define the selected offset
’s limits for Pass/Fail indication, complete the following steps:
a.
Press the
Lower
soft key to set the threshold for the power difference between the main
channel bandwidth and the selected offset bandwidth on the left.
b.
Enter a value by using the numeric keys. You can also use the rotary knob.
c.
Select the
dBc
soft key.
d.
Press the
Upper
soft key to set the threshold for the power difference between the main
channel bandwidth and the selected offset bandwidth on the right.
e.
Enter a value by using the numeric keys. You can also use the rotary knob.
f.
Select the
dBc
soft key.
Содержание CellAdvisor JD780A Series
Страница 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Страница 91: ...JD780A Series Analyzers Using Spectrum Analyzer 91 Figure 23 PIM detection for multiple carriers...
Страница 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Страница 525: ......