![JDS Uniphase CellAdvisor JD780A Series Скачать руководство пользователя страница 120](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616120.webp)
JD780A Series Analyzers
120
Using Interference Analyzer
a.
Press the
Edit Limit
soft key.
b.
Press the
Move
soft key and then turn the rotary knob to select the segment to edit.
c.
Select the menu option, from the following choices:
–
To hide the line for the selected segment, toggle the
Line
soft key and select
Off
.
–
To add a new point, press the
Add Point
soft key.
–
To delete the selected point, press the
Delete Point
soft key.
–
To change the position, press the
Frequency
or
Amplitude
soft key, and then turn the
rotary knob to change the value as desired.
To define the channel limit:
1.
Press the
LIMIT
hot key.
2.
Press the
Channel Limit
soft key.
3.
To define the index table, complete the following steps:
a.
Toggle the
Table
soft key between
On
and
Off
to display and hide the channel index table
on the screen.
b.
Press the
Index
soft key.
c.
Enter an index number to be set between 1 and 10 by using the numeric keys.
d.
Press the
Enter
soft key.
e.
Press the
Start Frequency
or
Center Frequency
soft key depending on your prior
frequency setting.
f.
Enter a value by using the numeric keys. You can also use the rotary knob or the arrow
keys.
g.
Select the unit:
GHz
,
MHz
,
kHz
, or
Hz
.
h.
Press the
Stop Frequency
or
Span Width
soft key depending on your prior frequency
setting.
i.
Enter a value by using the numeric keys. You can also use the rotary knob or the arrow
keys.
j.
Select the unit:
GHz
,
MHz
,
kHz
, or
Hz
.
4.
To set the limits for Pass/Fail indication, complete the following steps:
a.
Toggle the
Limit
soft key between
On
and
Off
to enable and disable the limit.
b.
Press the
High Limit
soft key.
c.
Enter a value by using the numeric keys. You can also use the rotary knob or the arrow
keys.
d.
Press the
dBm
soft key.
e.
Press the
Low Limit
soft key.
f.
Enter a value by using the numeric keys. You can also use the rotary knob or the arrow
keys.
g.
Press the
dBm
soft key.
5.
Optional
. Go to
SAVE/LOAD
> Save
, and then select
Limit
to save the limit settings.
See
“Using save” on page 31 for more information.
Содержание CellAdvisor JD780A Series
Страница 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Страница 91: ...JD780A Series Analyzers Using Spectrum Analyzer 91 Figure 23 PIM detection for multiple carriers...
Страница 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Страница 525: ......
Страница 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...