![JDS Uniphase CellAdvisor JD780A Series Скачать руководство пользователя страница 361](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616361.webp)
JD780A Series Analyzers
Using CDMA Signal Analyzer
361
Setting limit for OTA channel scanner
Procedure
1.
Press the
LIMIT
hot key.
2.
Press the
Limit Line
soft key to set a threshold for the limit line and Pass/Fail indication.
3.
Enter a value, and then press the
dBm
unit soft key.
4.
Toggle the
Limit Line
soft key between
On
and
Off
to display or dismiss the limit line.
5.
Optional
. You can enable alarm sound that goes off if the measurement falls outside of the limit.
Toggle the
Beep
soft key between
On
and
Off
to enable or disable the beep sound.
6.
Optional
. Go to
SAVE/LOAD
> Save
, and then select
Limit
to save the limit settings.
See
“Using save” on page 31 for more information.
Measurement example
Figure 170
Channel scanner measurement with CDMA OTA signal analyzer
PN scanner
Each access network sector has a unique PN offset. The PN Scanner is used to identify all the active
PNs in an area.
Setting measure setup
After configuring test parameters as described in the
“Configuring test parameters” on page 334, you
can continue your measurement. The measurement settings can be saved and recalled as a file. You
can also use JDViewer, PC application software to configure a measure setup, save as a file, and load
the file on to the instrument.
Procedure
1.
Press the
MEASURE SETUP
hot key.
2.
To set the incremental step for the PN offset, complete the following steps:
Содержание CellAdvisor JD780A Series
Страница 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Страница 91: ...JD780A Series Analyzers Using Spectrum Analyzer 91 Figure 23 PIM detection for multiple carriers...
Страница 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Страница 525: ......
Страница 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...