JD780A Series Analyzers
Using Interference Analyzer
101
–
To automatically set, toggle the
Sweep Time
soft key and select
Auto
.
–
To manually set, complete the following steps:
a.
Press the
Sweep Time
soft key.
b.
Enter a value by using the numeric keys.
c.
Select the unit:
Sec
,
mSec
, or
µSec
.
The sweep time mode automatically changes to
Manual
.
Sweep mode
The default settings of the sweep mode are
Continue
and
Normal
to sweep continuously at a normal
speed for most on-going measurements. If you want to hold the measurement or get a single sweep,
you can change the sweep mode.
Procedure
To select the single sweep mode:
1.
Press the
SWEEP
hot key.
2.
Toggle the
Sweep Mode
soft key and select
Single
. You can also use the
HOLD
hot key.
The letter
HOLD
in red appears and the sweeping is paused.
3.
Optional
. Press the
Sweep Once
soft key to get a new measurement.
To return to the continuous sweep mode:
1.
Toggle the
Sweep Mode
soft key and select
Continue
. You can also use the
HOLD
hot key.
The letter
HOLD
in red disappears and the sweeping resumes.
NOTE
Selecting
Sweep Mode > Fast
may reduce the measurement accuracy, but you can use
this fast sweep mode to identify the existence of interfering signals.
Gated sweep
You can configure the gated sweep settings with an internal or external trigger source. If you select
Gated Sweep Settings
, you can view the signal in time domain.
Procedure
1.
Press the
SWEEP
hot key.
2.
Press the
Gated Sweep Settings
soft key.
3.
Optional
. Adjust sweep time to allow enough time for a cycle by completing the following steps:
a.
Press the
Zero Span Time
soft key.
b.
Enter a value by using the numeric keys.
c.
Select the unit:
Sec
,
mSec
, or
µSec
.
4.
Set the period, if using the internal trigger source, by doing one of the following:
–
To select the preset period, complete the following steps:
a.
Toggle the
Period
soft key and select
StdSignal
.
b.
Press the
StdSignal
soft key, and then select the period:
GSM (4.615 ms)
,
WCDMA
(10 ms)
,
LTE (10 ms)
,
EV-DO (5 ms)
,
TD-SCDMA (5 ms)
, or
WiMAX (10 ms)
.
–
To manually set the period, complete the following steps:
Содержание CellAdvisor JD780A Series
Страница 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Страница 91: ...JD780A Series Analyzers Using Spectrum Analyzer 91 Figure 23 PIM detection for multiple carriers...
Страница 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Страница 525: ......
Страница 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...