JD780A Series Analyzers
Using LTE - FDD Signal Analyzer
243
2.
Toggle the
IQ Diagram
soft key between
On
and
Off
to display or dismiss data on the IQ
diagram for each component carrier.
3.
Press the
Channel
soft key to select the channel to be displayed on the IQ diagram.
4.
Use the rotary knob to move and select an item under EVM.
Conducting LTE - FDD OTA measurements
The FDD LTE Over The Air (OTA) measurement has Channel Scanner, ID Scanner, Multipath Profile,
Datagram, and Control Channel screens. The ID Scanner displays key parameters such as RSRP,
RSRQ, RS-SINR, and S-SS RSSI. The Multipath Profile enables you to determine RF environmental
conditions of testing area. The OTA Control Channel provides summary of all control channels
including RS power variation over time.
Channel scanner
The Channel Scanner is a radio receiver that can automatically tune or scan two or more discrete
frequencies and multi-channels, indicating when it finds a signal on one of them and then continuing
scanning when that frequency goes silent.
Setting channel/frequency
Procedure
To set the channels to be scanned:
1.
Press the
FREQ/DIST
hard key.
2.
Toggle the
Unit
soft key and select
Channel
.
3.
Press the
Index
soft key, and then enter an index number between one and six by turning the
rotary knob or using the numeric keys.
4.
To select the standard channel, complete the following steps:
a.
Press the
Channel Std
soft key. The standard channel window appears.
See
“Appendix C – Band, frequency & channel standard” on page 571 for more information.
b.
Select the band to be measured by using the rotary knob, the arrow keys, or the
Page
Up/Page Down
soft keys.
c.
Press the
Select
soft key or the rotary knob to confirm the selection.
5.
To set the channel number for the selected index, complete the following steps:
a.
Press the
Channel Number
soft key.
b.
Enter a value by using the numeric keys. You can also use the rotary knob.
c.
Press the
Enter
soft key.
The instrument displays a corresponding center frequency for the channel number.
6.
To set the integration bandwidth for the selected index, complete the following steps:
a.
Press the
Integration Bandwidth
soft key.
b.
Enter a value by using the numeric keys. You can also use the rotary knob.
c.
Select the unit:
GHz
,
MHz
,
kHz
, or
Hz
.
7.
To add more channels to be scanned, repeat steps
3-6
.
8.
Press the
ESC
hard key to dismiss the channel list window and view the scanning result.
Содержание CellAdvisor JD780A Series
Страница 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Страница 91: ...JD780A Series Analyzers Using Spectrum Analyzer 91 Figure 23 PIM detection for multiple carriers...
Страница 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Страница 525: ......
Страница 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...