JD780A Series Analyzers
496
Using Cable and Antenna Analyzer
The calibration progress bar appears.
7.
Connect the
LOAD
connector of the CAL Kit directly to the
Cable and Antenna Analyzer
Reflection/RF Out
port or at the end of the connected extension cable.
8.
Press the
Continue
soft key to continue calibration.
The calibration progress bar appears.
After completion, the calibration status on the screen changes to
ON
.
O-S-L calibration using electronic EZ-Cal kit
The e-Calibration is an Open-Short-Load calibration using the electronic EZ-Cal kit, which lets you
perform the O-S-L calibration easy and fast with only one connection of the kit to the instrument.
The O-S-L calibration must be performed after setting the frequencies and connecting an extension
cable as like any mechanical calibration.
illustrates the connection when a port extension
cable is used for calibration.
Figure 247
1-port calibration connection with electronic EZ-Cal kit
Procedure
1.
Connect an extension cable, if necessary, to the
Reflection/RF Out
port of the instrument.
2.
Connect your
EZ-Cal
to the other open end of the extension cable or to your instrument's
Reflection/RF Out
port directly.
3.
Connect the USB cable to the side of the
EZ-Cal
and then to the
USB Host
port to power the
e-calibration kit.
The instrument detects the EZ-Cal automatically and displays the EZ-Cal icon.
4.
Press the
MEASURE SETUP
hot key.
5.
Press the
Calibrate
soft key.
The on-screen instruction appears. to guide you through the e-calibration.
6.
Press the
Continue
soft key to continue calibration.
The calibration progress bar appears.
After completion, the calibration status on the screen changes to
ON (E)
.
Quick calibration in DTF
The Quick calibration that is available only in the DTF measurement mode is useful when you want to
measure the cable length only, without having to perform a full O-S-L calibration with a mechanical
Y-Cal kit.
Содержание CellAdvisor JD780A Series
Страница 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Страница 91: ...JD780A Series Analyzers Using Spectrum Analyzer 91 Figure 23 PIM detection for multiple carriers...
Страница 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Страница 525: ......
Страница 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...