JD780A Series Analyzers
Using Interference Analyzer
107
Dual spectrogram
The Dual Spectrogram mode is useful to compare two spectrograms displayed simultaneously on the
screen.
Setting measure setup
Procedure
1.
Configure test parameters as described in the
“Configuring test parameters” on page 95.
2.
Press the
MEASURE SETUP
hot key.
3.
Toggle the
Active Window
soft key and select
Left
or
Right
spectrogram to be set up.
4.
To set the amount of time between each trace measurement:
a.
Press the
Time Interval
soft key.
b.
Enter a value by using the numeric keys. You can also use the rotary knob.
c.
Press the
Sec
soft key.
5.
To set the time cursor on a specific trace position, complete the following steps:
a.
Toggle the
Time Cursor
soft key and select
On
.
b.
Turn the rotary knob to move the time cursor. You can also enter a value and then press the
Enter
soft key.
NOTE
Enabling the time cursor puts the measurement on hold and you can make post-
processing analysis for each measurement over time using the time cursor.
6.
To start a new measurement, press the
Reset/Restart
soft key.
7.
Optional
. Go to
SAVE/LOAD
> Save
, and then select the save option from the choices available
for your measurement mode. See
“Using save” on page 31 for more information.
Measurement example
Figure 29
Interference measurement in dual spectrogram view
Содержание CellAdvisor JD780A Series
Страница 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Страница 91: ...JD780A Series Analyzers Using Spectrum Analyzer 91 Figure 23 PIM detection for multiple carriers...
Страница 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Страница 525: ......
Страница 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...