JD780A Series Analyzers
Using Mobile WiMAX Signal Analyzer
459
External offset
You can turn the
External Offset
on and manually set the external offset value. An offset consists of a
cable loss and a user offset and the measurement result shows the value reflecting both offset values.
When the external offset value is set at 40 dB in the Spectrum mode, the measurement result
compensates 40 dB at both the Spectrum Analyzer and Signal Analyzer modes
Procedure
To set the external offset:
1.
Press the
AMP/SCALE
hard key.
2.
Toggle the
External Offset
soft key and select
On
.
3.
Enter a value by using the numeric keys.
4.
Press the
dB
soft key to complete the entry.
To turn the external offset off:
1.
Press the
AMP/SCALE
hard key.
2.
Toggle the
External Offset
soft key and select
Off
.
Setting average
You can set the number of measurements to be averaged for the trace presentation. A maximum of
100 times of averaging can be set. When the averaging reaches to your setting, a new measurement
value replaces the measurement value in sequence from the earliest.
Procedure
1.
Press the
BW/AVG
hard key.
2.
Press the
Average
soft key.
3.
Enter a value between 1 and 100 as needed by using the numeric keys.
4.
Press the
Enter
soft key.
Setting sweep mode
The default setting is
Continue
to sweep continuously for most on-going measurements. If you want to
hold the measurement or get a single sweep, you can change the sweep mode.
Procedure
To select the single sweep mode:
1.
Press the
SWEEP
hot key.
2.
Toggle the
Sweep Mode
soft key and select
Single
. You can also use the
HOLD
hot key.
The letter
HOLD
in red appears and the sweeping is paused.
3.
Optional
. Press the
Sweep Once
soft key to get a new measurement.
To return to the continuous sweep mode:
1.
Toggle the
Sweep Mode
soft key and select
Continue
. You can also use the
HOLD
hot key.
The letter
HOLD
in red disappears and the sweeping resumes.
Содержание CellAdvisor JD780A Series
Страница 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Страница 91: ...JD780A Series Analyzers Using Spectrum Analyzer 91 Figure 23 PIM detection for multiple carriers...
Страница 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Страница 525: ......
Страница 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...