![JDS Uniphase CellAdvisor JD780A Series Скачать руководство пользователя страница 515](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616515.webp)
JD780A Series Analyzers
Using Cable and Antenna Analyzer
515
4.
To clear, write or capture, or hide the trace(s), do one of the following:
To
Select
Trace Legend
Clear current data and display with new
measurements
Clear Write
W
Capture the selected trace and compare traces
Capture
C
Load a saved trace
Load
L
Hide the selected trace
Trace View > Off
F
Remove all the traces and initialize the trace
settings
Trace Clear All
5.
Optional.
If you have at least two traces (T1 and T2), you can perform trace math:
a.
Press the
Trace Math
soft key.
b.
Press the
T1
– T2 -> T5
or
T2
– T1 -> T6
soft key.
NOTE
The trace to be loaded must be in the same measurement mode and has the same
frequency setting as the current measurement.
Setting zoom zones
You can define up to three zoom-in zones so that you can view uplink and downlink frequencies in
detail on a single measurement window for compliance verification.
Procedure
1.
Press the
MEASURE SETUP
hot key.
2.
Press the
Zoom
soft key.
3.
To view a zoom zone in a separate chart, complete the following steps:
a.
Press the
Zoom
soft key, and then select the zone number to zoom in:
Zone 1
,
Zone 2
,
or
Zone 3
.
b.
Toggle the
Zoom
soft key and select
On
.
The zoom-in chart for the selected zone appears under the measurement result.
4.
To define a zone range, complete the following steps:
a.
Press the
Zone [1|2|3]
soft key to select an active zone.
b.
Toggle the
Zone [1|2|3]
soft key between
On
and
Off
to view or dismiss the active zone.
c.
Press the
Start Frequency
soft key.
d.
Enter a value by using the numeric keys. You can also use the rotary knob.
e.
Select the unit:
GHz
,
MHz
,
kHz
, or
Hz
.
f.
Press the
Stop Frequency
soft key.
g.
Enter a value by using the numeric keys. You can also use the rotary knob.
h.
Select the unit:
GHz
,
MHz
,
kHz
, or
Hz
.
5.
To clear all the zones, press the
Clear All
soft key.
Setting alternate DTF band
You can use the Alt DTF Band in DTF measurements to scale down a specific sub-band without a
Содержание CellAdvisor JD780A Series
Страница 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Страница 91: ...JD780A Series Analyzers Using Spectrum Analyzer 91 Figure 23 PIM detection for multiple carriers...
Страница 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Страница 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Страница 525: ......
Страница 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...