Follow the directions for positioning of the cube in the loader.
Do not remove the protective strip before installing the cube in the loader.
Install only full cubes. Never install individual plates.
Reassembly of cuvette plates will cause the analyser to jam.
Correct positioning of the cube of cuvettes in the loader is essential for the automatic cuvette
loading module to function correctly.
The cuvettes are disposable devices.
!
Refer to the catalogue for the product references of the ancillary reagents to be ordered.
Use
3-3- Installation of Ancillary Reagents (continued)
IDS-iSYS User Manual - Revision M1
Use 3-3
51