3-2-1- Colour codes associated with reagent positions in the reagent compartment
Rail free.
Reagent correct.
???
Presence of a rack or reagent detected but not identified.
Preventive alarm threshold reached.
Reagent volume inadequate or deadline for using over.
Information associated with the reagent not available.
Reagent detected higher than the maximum filling level.
3-2-2- Information displayed with reagents
The main information regarding on-board reagents is directly displayed via its graphic representation.
Detailed information is displayed by clicking on the relevant reagent.
Immunoassay cartridges
Deadline for using the cartridge on the system. Date is replaced by expiry time
when the day of expiry date begins.
Countdown before use (after reagent loading).
Test name and lot number.
Number of remaining tests.
Number of available dilutions (only for tests using a diluent).
A calibration will be requested in the next 24 hours for tests using a calibration
frequency managed by the system.
The calibration was recalculated after excluding at least one calibrator replicate
Status of the current calibration for the reagent lot (see page 44).
Use
3-2- Installation of Reagents (continued)
IDS-iSYS User Manual - Revision M1
Use 3-2
43
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...