A6-2-Between 15 and 30 days
The analyser can be completely stopped after a decontamination of the IDS-iSYS TRIGGER B fluidic circuit
followed by the emptying of the two TRIGGER and of the D-SORB fluidic circuits. The TRIGGER tubing
cleaning maintenance, included in the weekly maintenance, will be used to carry out this procedure.
•
Remove all samples.
•
Remove reagent racks and store in accordance with IFU recommendations.
•
Carry out daily maintenance.
•
Carry out the general steps of the weekly maintenance.
•
Perform the «Rinsing of the Trigger A and Trigger B fluid lines» maintenance a first time for rinsing the fluidic
circuits:
•
Replace the two Trigger bottles by
two 250 mL bottles containing distilled water
.
•
When the first step of the maintenance is completed
, continue the maintenance with the two bottles
containing distilled water.
•
When the rinsing is over, perform the «Rinsing of the Trigger A and Trigger B fluid lines» maintenance a
second time for decontamination of the fluidic circuits:
•
Keep
the distilled water bottle in the TRIGGER A position
. Install
the
TRIGGER A bottle at the
TRIGGER B position
.
•
When the first step of the maintenance is completed
, continue the maintenance with the same bottles
installed.
•
When the decontamination is over, perform the «Rinsing of the Trigger A and Trigger B fluid lines»
maintenance a third time for rinsing and emptying the fluidic circuits:
•
Replace the two Trigger bottles by
two 250 mL bottles containing distilled water
.
•
When the first step of the maintenance is completed
, replace the two bottles containing distilled water
with
empty bottles
. Continue the maintenance.
•
Carry out the general steps of the monthly maintenance.
•
For systems using AP Substrate, perform the « Rinsing of the AP Substrate tubing» maintenance twice to
rinse the fluidic circuit following the below steps:
•
Replace the AP Substrate bottle by a 500 mL bottle containing de-ionised water to mimic the AP
Substrate.
•
Click on the corresponding
button in front of this item in the maintenance list.
•
Click on
.When the first step is completed, a message is displayed.
•
Keep
the distilled water bottle in the AP Substrate position
, continue the maintenance by clicking on
.
•
Repeat this procedure a second time.
•
After the two rinsing procedures are completed, perform the « Rinsing of the AP Substrate tubing»
maintenance for a third time to complete the rinse and to empty the fluidic circuit:
•
Keep
the distilled water bottle in the AP Substrate position
.
•
Click on the corresponding
button in front of this item in the maintenance list.
•
Click on
.When the first step is completed, a message is displayed.
•
Replace the bottle containing distilled water with an
empty bottle
. Continue the maintenance by
clicking on
.
•
Remove the bottle in the AP Substrate position.
Appendices
IDS-iSYS User Manual - Revision M1
Appendix A-6
A 11
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...