ASTM Name of the field
Value*
Description and use in IDS-iSYS system
Priority
R
This field may contain the values:
•
S
= STAT.
•
A = ASAP.
•
R
= Routine.
•
C = Callback.
•
P = Preoperative.
For the device, only S has any influence, the
passage of the associated profile into Urgent, and
the other values are ignored.
Requested/Ordered
Date and Time
Not used.
Specimen Collection
Date and Time
Not used.
Collection End Time
Not used.
Collection Volume
Not used.
Collector ID
Not used.
Action Code
A
Action to be carried out on the block test or tests.
This field may contain the values:
•
A
for Add.
•
C for Cancel.
•
N
for New.
•
P for Pending.
•
Q for QC.
For the device, A and N are managed in the same
way (new request or new rerun), C (cancellation of
a request) can be managed if the test is not in
progress. P and Q are not managed.
Danger Code
Not used.
Relevant Clinical Information
Not used.
Date and Time
Specimen Received
Not used.
3- ASTM Compatible V2 Version
3-4- The Order Block (continued)
IDS-iSYS Connection Protocol - Revision N1
3- ASTM Compatible V2 Version
25
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...