3-4- System Performance Checks (Immunoassay only)
The performance of the analyser must be checked on a daily basis before performing an assay requiring
calibration, controls or samples.
To achieve this, a reagent cartridge known as Cartridge Check System (CCS) is used. Di
ff
erent protocols
are applied to this cartridge in order to determine the functional state of the various analyser modules.
For daily use and acceptance criteria, refer to the CCS cartridge instructions for use (IFU).
The qualification profile can be programmed as part of the automatic start-up of the analyser (see Section 7,
7-1- Programming the automatic qualification profile
, page 152).
3-4-1- Programming the qualification profile
•
From the main screen, select:
The qualification profile menu is displayed.
•
If results from a previous profile are present in the list of current results, click on
.
The results are sent to storage and the list is released for a new profile.
•
From the drop-down list, select the necessary test and programme the required number of replicates (see
CCS IFU) then click on
.
•
Repeat the operation for each test required.
•
Click on
.
•
Start the profile by clicking on
.
Use
3-4- System Performance Checks (Immunoassay only)
IDS-iSYS User Manual - Revision M1
Use 3-4
52
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...