Issues in software package V 14.04 (continued)
•
In the Storage menu displayed in standard mode, when the transfer to the LIS of a patient profile is requested while
one of multiple replicate results is selected, this selected result is not sent with the profile results. The relevant profile
must be selected to ensure sending of the complete profile.
•
In certain cases, assays schedule to a first cartridge but not complete, may remain in error (red tick) despite a second
cartridge of the same lot being on-board. The first cartridge may also not be fully used. In this case, the run cycle must
be stopped and the relevant assays must be deselected (uncheck the box) prior to be selected again.
•
Rounding at 0.1 mAbs may be di
ff
erent for the same result.
•
When two analytes use the same cartridge, the number of available tests can be incorrect. In this case, all assays
requested for the two analytes will be carried out.
•
The blue contour is applied only to removable racks in the sample compartment.
•
A countdown before use may not resume when the cartridge is removed less than 5 minutes. In this case, a new
countdown begins.
•
The status of a faulty calibration is not updated when excluding one of the calibrator replicates during a run. Patient
results and controls are calculated. The calibration status will be updated once the run cycle is stopped.
Preface
IDS-iSYS User Manual - Revision M1
P 4
Software version V 14
Содержание IDS-iSYS
Страница 7: ......
Страница 13: ...SECTION 1 Operating principle...
Страница 30: ...SECTION 2 User Interface Software...
Страница 48: ...SECTION 3 Use...
Страница 119: ...SECTION 4 Messages...
Страница 130: ...SECTION 5 Maintenance...
Страница 165: ...SECTION 6 Problems Corrective Action...
Страница 169: ...SECTION 7 System Configuration...
Страница 182: ...Appendices...