Excluding one of the replicates
For Immunoassay, individual outlying calibrator or control replicates can be excluded when the CV is above the
acceptable limit.
•
Click on
to display the measurements (RLU or mAbs.).
•
To exclude one of the measurements, click on the corresponding red tick.
•
The calibration and the control(s) are calculated with the new curve equation.
•
When points have been excluded, the icon
is displayed with the calibration date.
•
When the recalculated calibration is valid, with the control(s) within acceptable limits, the message
OK
is
displayed.
•
Click on
to save the modification.
•
If the calibration is not validated, the
button is displayed.
•
In this case, request a new calibration after having replaced failed calibrator/control.
Use
3-10-
Results (continued)
IDS-iSYS User Manual - Revision M1
Use 3-10
84